Skip to main content
. Author manuscript; available in PMC: 2018 Aug 7.
Published in final edited form as: Phys Med Biol. 2017 Aug 7;62(17):6938–6962. doi: 10.1088/1361-6560/aa7ccc

Table 7.

Experimental and simulated spatial average of the delay (δ̂ and δ) and transmitted energy ( and P) for specimen IV. The spatial averages of δ̂ and δ were calculated on the XY region showing normalized pressure greater than 0.5 in water-only conditions. The transmitted powers and P were calculated over the whole XY plane of the measurements. |Δ| indicates the absolute of the difference.

270 kHz
Specimen Loc. δ̂
(µs)
δ
(µs)
|Δ|
(µs)

%
P
%
|Δ|
%
Ξ
IV 1 −1.29 −0.36 0.93 22 3.7 18 0.32
2 1.88 0.82 1.06 12 3.6 8.6 0.32
3 1.53 1.85 0.32 7.3 7.4 0.15 0.18

Mean (±s.d.) 0.77(±0.32) 9.1(±7.5) 0.27(±0.07)
836 kHz
Specimen Loc. δ̂
(µs)
δ
(µs)
|Δ|
(µs)

%
P
%
|Δ|
%
Ξ
IV 1 0.69 0.23 0.46 4.4 0.8 3.6 0.29
2 0.69 0.22 0.47 4.2 0.4 3.8 0.44
3 0.94 −0.04 1.00 0.07 0.01 0.02 0.48

Mean (±s.d.) 0.65(±0.27) 2.5(±2) 0.4(±0.08)