TABLE 2.

Intraoperative and Immediate Postoperative Results

Attribute Outcome
Umbilical incision size (mm)—average 10.1
Grasper application times (sec)
 Time between grasper introduction and correct position and release—average (SD) 23 (19)
 Time between correct position and coupling with external magnet—average (SD) 50 (60)
Procedure times (min)
 Overall procedure time—average (SD) 63 (19)
Grasper times (min)
 Time magnetic grasper in abdomen—average (SD) 36 (0:12)
 Time magnetic grasper coupled with magnetic controller—average (SD) 34 (0:12)
Device ease of use
 Yes 50
 No 0
Quality of exposure
 Excellent 45
 Sufficient 5
 Insufficient 0
Device malfunctions
 Yes 0
 No 50
Conversions (%)
 Open surgery 0
Four (4)-port surgery 1*
Estimated blood loss (mL)
 Number of subjects with blood loss 32
 Average within all subjects (SD), n = 50 7 (11)
 Average within subjects with blood loss (SD), n = 32 11 (11)
Estimated bile spilled (mL)
 Number of subjects with bile spilled 6
 Average within all subjects (SD), n = 50 5 (18)
 Average within subjects with bile spilled (SD), n = 6 39 (41)
PACU Time (min)—average (SD) 137 (58)
Pain scores
 0 Hours postprocedure—average (min, max), n = 48 1.0 (0, 6)
 3 Hours postprocedure—average (min, max), n = 50 1.8 (0, 8)
 6 Hours postprocedure—average (min, max), n = 50 1.5 (0, 6)
 At discharge—average (min, max), n = 50 1.3 (0, 4)
External abdominal wall evaluation (at 6 hours postop)
 Normal 50
 Redness 0
 Bruising 0
 Other 0

*Converted after the use of device was complete.

PACU indicates postanesthesia care unit; SD, standard deviation.