Skip to main content
. 2018 Jan 4;13(1):e0187663. doi: 10.1371/journal.pone.0187663

Table 2. Blood hemoglobin concentrations of participant groups as determined by various methods.

Parameter Testing method
Sysmex KN21TM HemoCue® wicking Sysmex KN21TM HemoCue® gravity Sysmex KN21TM Pronto® DCI-mini™
LabHb
(n = 66)
WHb
(n = 66)
LabHb
(n = 66)
GHb
(n = 66)
LabHb
(n = 112)
PrHb
(n = 112)
Hb Mean (SD) 12.0 (1.4) 11.6 (1.5) 11.7 (1.5) 11.8 (1.7) 11.9 (1.2) 11.6 (0.9)
Hb Range 8.4–16.2 7.9–16.1 7.3–19.5 6.5–19.4 7.3–15.0 9.7–14.3
% Anemic 36% 44% 38% 39% 34% 42%
Pearson correlation WHb vs LabHb GHb vs LabHb PrHb vs LabHb
Estimate 0.8* 0.9* 0.43*
Bias WHb-LabHb GHb-LabHb PrHb-LabHb
Mean (SD) -0.3 (1.0) 0.0 (0.7) -0.2 (1.1)
95% CI -0.6 –-0.1 -0.1–0.2 -0.5–0.0
Range -3.1–1.8 -2.9–1.7 -3.7–2.6
95% Limits of Agreement*
-2.3–1.7 -1.4–1.4 -2.4–2.0
Absolute Difference (<1.0g/dl)
Mean (SD) 0.7 (0.7) 0.4 (0.5) 0.9 (0.7)
Range 0–3.1 0–2.9 0–3.7

* 95% Limits of Agreements were calculated as the mean bias ± 1.96 x SD