Table 4. Factors influencing the ability of the Pronto® device to generate readings.
Able to detect reading? | |||
---|---|---|---|
Yes | No | p-value | |
Mean (SD) | Mean (SD) | ||
(n = 112) | (n = 20) | ||
Age, year | 23.3 (15.1) | 21.5 (15.1) | 0.62 |
Child birth rank | 2 (2) | 2 (1) | 0.90 |
Height | 83.1 (13.4) | 78.3 (12.6) | 0.14 |
HAZ | -0.4 (1.6) | -1.2 (1.7) | 0.05 |
WHZ | -0.8 (1.7) | -0.8 (1.8) | 0.95 |
MUAC | 14.8 (1.7) | 13.2 (2.4) | 0.00 |
Skin pigmentation | 6 (1) | 6 (1) | 0.68 |
Pulse rate | 133 (19) | 129 (28) | 0.37 |
SpO2% | 96.7 (1.7)% | 92.5 (8.2) % | 0.00 |
Perfusion Index | 4.0 (2.0) | 3.4 (2.0) | 0.20 |
Hemoglobin | 11.9 (1.2) | 11.6 (2.7) | 0.45 |