. 2017 Oct 24;46(1):97–107. doi: 10.1007/s10439-017-1941-2

Table 1.

Performance analysis of implant loosening detection built upon, input–output coherence correlation and different output spectrum features.

Coherence correlation (%) New peak in (750–900 Hz) (%) Peak shift in (700–1200 Hz) (%) Peak shift in (1200–2200 Hz) (%) Peak flattening (%)
Sensitivity 91.67 ± 0.03 77.2 ± 6.8 53.1 ± 13.7 66.2 ± 9.0 51.0 ± 4.6
Specificity 92.86 ± 0.00 78.9 ± 3.9 74.3 ± 10.5 43.9 ± 5.8 84.6 ± 10.2
Accuracy 92.26 ± 0.01 78.0 ± 3.5 63.0 ± 5.8 57.2 ± 8.1 67.3 ± 6.0

The mean and standard deviation of each metric were obtained over repeated measurements over 14 legs