Published in final edited form as: Magn Reson Med. 2017 Jul 7;79(3):1495–1505. doi: 10.1002/mrm.26819

Table 1.

Image quality scores (average ± standard deviation) from two musculoskeletal radiologists for Standard MSI and accelerated MSI with bin-by-bin compressed sensing and RPCA reconstructions. A one-way paired Wilcoxon test was used to assess the null hypothesis that RPCA and bin-by-bin CS images were at least one point worse than standard MSI images. All null hypotheses were rejected at the 0.05 level.

Standard MSI Bin-by-bin CS RPCA
Mean ± Std. Dev. (Reader 1/Reader 2) Mean ± Std. Dev. (Reader 1/Reader 2) P-value Mean ± Std. Dev. (Reader 1/Reader 2) P-value
Image quality 3.5 ± 0.71/3.5 ± 0.50 3.1 ± 0.60/3.1 ± 0.60 0.017 3.2 ± 0.43/3.4 ± 0.48 0.004
Sharpness away from metal 3.6 ± 0.48/3.5 ± 0.50 3.1 ± 0.33/3.0 ± 0.50 0.007 3.5 ± 0.50/3.5 ± 0.50 0.001
Sharpness near metal 3.4 ± 0.48/3.1 ± 0.78 3.0 ± 0.50/3.0 ± 0.71 < 0.001 3.2 ± 0.43/3.0 ± 0.71 0.001
Blockiness/artificial appearance 3.9 ± 0.33/3.8 ± 0.43 3.5 ± 0.71/3.5 ± 0.71 0.009 4.0 ± 0.00/4.0 ± 0.00 < 0.001
Signal loss near metal 3.2 ± 0.66/3.0 ± 0.87 3.4 ± 0.48/3.0 ± 0.71 < 0.001 3.6 ± 0.48/3.4 ± 0.70 < 0.001