Skip to main content
. 2012 Apr 24;2(5):390–393. doi: 10.1016/j.jpha.2012.04.006

Table 3.

Results of accuracy and precision studies for eplerenone.

Level Concentration added (ng/mL) Intra-day (n=5)
Inter-day (n=5)
Recovery (n=5)
Meanconcentration found (ng/mL) RE (%) RSD (%) Meanconcentration found (ng/mL) RE (%) RSD (%)
LQC 250 255.87 2.35 6.28 255.58 2.23 5.45 80.75
MQC 1000 947.48 −5.25 8.97 976.19 −2.38 5.80 81.23
HQC 3000 3141.81 4.73 6.01 3070.23 2.34 5.79 74.35
IS 82.20