Skip to main content
. 2013 Apr 29;3(6):481–488. doi: 10.1016/j.jpha.2013.03.009

Table 3.

Stability of MC in plasma samples.

Stability Spiked plasma concentration (ng/mL) Concentration measured (ng/mL) (mean±SD; n=6) RSDa (%) (n=6)
Room temperature stability (24.5 h) 3.00 2.85±0.26 9.14
300.00 298.07±13.36 4.48
Processed sample stability (71.5 h) 3.00 3.08±0.09 2.84
300.00 310.46±8.09 2.60
Long term stability (55 days) 3.00 3.03±0.09 2.95
300.00 295.41±3.38 1.15
Freeze–thaw stability (cycle 3, 48 h) 3.00 2.86±0.20 7.10
300.00 297.11±7.46 2.51
a

(Standard deviation/mean concentration measured)×100.