Table 2.
Analysts | Actual concentration of sample (mg/L) | Standard addeda (mg/L) | Mean recoveryb (%)±SD | LOD (mg/L) |
---|---|---|---|---|
Creatinine | 861.25 | 160 | 97.90±0.84 | 0.010 |
320 | 93.49±0.79 | |||
Uric acid | 411.78 | 80 | 96.45±4.20 | 0.025 |
160 | 95.38±1.76 | |||
Hypoxanthine | 51.96 | 50 | 112.46±1.08 | 0.050 |
100 | 115.78±0.94 | |||
Xanthine | 54.44 | 20 | 97.13±3.16 | 0.025 |
50 | 90.82±0.96 |
Concentrations of standards are expressed as the equivalent concentrations added in the final injected solutions. Each level of standard added was replicated six times.
Calculated recovery (%)=(amount observed−original amount)/added amount×100%.