Skip to main content
. 2013 Aug 14;4(2):125–131. doi: 10.1016/j.jpha.2013.08.001

Table 1.

Analytical results obtained for determining AMK base content in AMK sulfate injections by the proposed CFS-FTIR-DS method.

Band peak parameters (units in cm−1) of the first derivative spectra
Parameters derived from linear regressiona: Y=a+bX
Measurement mode Location Baseline a±SD (×10−3) b±SD (×10−3) r LOD RSD
Total height 1514 Zero abs 0.08±0.02 0.0648±0.0004 0.9996 2.1 0.004
Corrected height 1514 1538–1469 0.05±0.07 0.0728±0.0007 0.9995 2.2 0.005
Total area 1514 Zero abs 1±1 2.25±0.02 0.9987 3.5 0.30
Corrected area 1514 1538–1469 1.0±0.6 2.81±0.01 0.9997 1.6 0.15
a

Calibration curves obtained using eight standards of AMK sulfate, in triplicate, registered as its equivalent AMK concentration ranging from 7.7 to 77.0 (mg/mL); where X is the explanatory variable and Y is the dependent variable; the slope of the line is b, and a is the Y-intercept; SD=standard deviation; r=correlation coefficient; LOD=limit of detection in mg/mL AMK base calculated as 3σ(1/slope); RSD=relative standard deviation (%).