. 2018 Jan 10;13(1):e0190585. doi: 10.1371/journal.pone.0190585

Table 3. Agreement between iSense and EVA scanners.

Comparing the absolute and relative means of the three iSense scan measurements (N = 14) and means of the three EVA scan measurements (N = 14), limits of agreement, coefficient of repeatability within methods and uniformity of the data for each section of the lower limb going from 300mm below the mid-patella (0mm) to 180mm above the mid-patella.

iSense vs. EVA
Sections (mm) Aver. circ. in mm Abs. mean diff. in mm (SE) Rel. mean diff. in % (SE) 95% Limit of Agreement (Mean ± 1.96SD) Repeatability Manual/EVA
Abs. Lower − Abs. Upper (mm) Rel. Lower − Rel. Upper (%) Abs. SE (mm)/Rel. SE (%)
180 517.0 -4.2 (2.6) -0.8 (0.5) -22.5 − 14.1 -4.5 − 2.8 18.8/0.7 6.5/6.0
150 493.1 -1.9 (3.1) -0.4 (0.6) -24.6 − 20.9 -5.1 − 4.3 27.8/1.2 7.2/5.3
120 464.2 0.3 (2.9) 0.0 (0.7) -21.3 − 21.8 -4.7 − 4.8 24.4/1.2 9.1/4.8
90 434.6 1.7 (2.3) -0.4 (0.5) -18.3 − 15.0 -4.3 − 3.5 13.9/0.8 8.5/5.3
60 411.5 -4.0 (1.8) -1.1 (0.4) -16.9 − 8.9 -4.1 − 2.2 7.8/0.5 5.4/7.1
30 398.3 -7.6 (1.7) -1.9 (0.4) -19.7 − 4.5 -5.0 − 1.2 6.7/0.4 7.6/5.6
0 386.1 -4.3 (1.5) -1.1 (0.4) -15.4 − 6.8 -4.0 − 1.8 5.2/0.4 8.3/5.9
-30 363.0 -2.7 (1.4) -0.8 (0.4) -12.7 − 7.3 -3.7 − 2.1 4.8/0.4 5.7/3.6
-60 353.1 -6.8 (1.8) -1.9 (0.5) -19.6 − 6.1 -5.5 − 1.6 8.0/0.6 8.4/3.0
-90 364.2 -8.3 (1.5) -2.3 (0.4) -19.4 − 2.8 -5.3 − 0.8 5.4/0.4 9.3/3.5
-120 377.6 -7.9 (1.2) -2.1 (0.3) -16.9 − 1.2 -4.4 − 0.2 3.7/0.2 6.3/3.8
-150 382.1 -3.6 (1.3) -0.9 (0.3) -12.8 − 5.7 -3.4 − 1.5 3.6/0.3 7.9/3.8
-180 371.2 0.2 (1.8) 0.0 (0.5) -13.0 − 13.3 -3.6 − 3.6 8.4/0.6 7.2/4.7
-210 347.2 3.4 (2.4) 1.0 (0.7) -13.9 − 20.7 -4.0 − 5.9 15.5/1.3 7.2/5.2
-240 316.9 2.8 (2.9) 0.9 (0.9) -18.5 − 24.1 -5.9 − 7.6 24.1/2.4 7.7/5.3
-270 287.8 2.9 (3.1) 1.0 (1.1) -19.7 − 25.5 -6.7 − 8.7 27.6/3.2 6.9/5.0
-300 268.1 2.1 (2.5) 0.9 (0.9) -14.1 − 18.4 -5.2 − 7.0 16.4/2.4 10.2/4.8

Aver., average; circ., circumference; abs., absolute; diff., difference; SE, 95% confidence interval; rel., relative; iSS, iSense. A positive bias or limits of agreement indicates that measurements from iSense scanner overestimates the value in respect to EVA scanner.