Table 1.

Quality Assurance Summary From In Vivo EPID Dosimetry on 10 Patients With Rectal Cancer.a

Patient No. Imaged Fields (Total: 164) Soft Tissue Changes in Belly Board Opening Belly Board Positioning Errors ≥ 1.5 cm Bowel Gas Patient Setup Errors ≥ 1 cm (Bony Anatomy) Other
1 8 (4 fields, 2 fx)
2 17 (3 fields, 6b fx) Board misaligned by 2 cm S-I (AP field, fx 5) Persistent gas (LAT fields, fx 2, 5, 6; Figure 4) 1 cm S-I setup error (AP field, fx 4)
3 12 (2 fields, 6 fx) Stretched soft tissue caused +8% AP field dose (both fields, all fx; Figure 2) Board misaligned by 2 cm S-I (both fields, fx 2, 5)
4 20 (4 fields, 5 fx) Stretched soft tissue caused +4% AP field dose (AP field, all fx) Tissue reduction in posterior region (LAT fields, all fx; Figure 5). Unexplained +10%-15% dose areas (PA field, all fx)
5 12 (2 fields, 6 fx) Board misaligned by 2 cm S-I (both fields, fx 1)
6 16 (4 fields, 4 fx) Board misaligned by 2 cm S-I (AP, PA fields, fx 3; Figure 3) 1.5 cm L-R setup error (AP field, fx 2), probable VRT setup error (LAT fields, fx 2)
7 21 (3 fields, 7 fx) Stretched soft tissue caused +3% AP field dose (AP field, all fx)
8 16 (4 fields, 4 fx)
9 18 (3 fields, 6 fx)
10 24 (3 fields, 8 fx) Genitals placed differently (AP field, all fx)

Abbreviations: AP, antero-posterior; fx, fraction(s); LAT, lateral; L-R, left-right; PA, postero-anterior; S-I, superior-inferior; VRT, vertical.

aIn brackets are the imaged fields and fractions in which the dose discrepancy was visible (columns 3-7).

bAntero-posterior field fraction 6 is not usable due to imager misplacement.