. 2015 Apr 24;5(6):356–370. doi: 10.1016/j.jpha.2015.04.002

Table 6.

Accuracy and precision studies of the HPLC method for determination of Eze-1 related substances.

Compound Accuracy (n=3)
Precision
Intra-day (n=6)
Inter-day (n=6)
Amount in test soln. (%) Amount theo. (Mean, %) Amount found (Mean±SD, %) RSD (%) Recovery (%) Amount found (Mean±SD, %) RSD (%) Amount found (Mean±SD, %) RSD (%)
4-Hydroxy benzaldehyde 0.094 0.242 0.245±0.008 3.2 100.9 0.10±0.004 4.0 0.10±0.004 4.0
0.492 0.495±0.009 1.8 100.5
0.735 0.740±0.009 1.2 100.7
4-Fluoroaniline 0.014 0.261 0.263±0.003 1.2 100.9 0.02±0.00 0.0 0.02±0.00 0.0
0.513 0.521±0.007 1.3 101.5
0.756 0.787±0.031 3.9 104.2
Desfluoro Eze-1 0.077 0.077a 0.076±0.001 1.5 NA 0.08±0.00 0.0 0.08±0.00 0.0
0.097 0.096±0.0006 0.60 98.63
0.146 0.149±0.001 0.67 102.1
a

Not spiked. NA: not applicable.