Table 3.

Relative linear stopping power comparisons for several Gammex tissue characterization plugs using Protocol

Gammex Material RSLP ‐100 MeV
Name Avg HU ρ [g/cm−3] RayStationa Stoich.b Fippelc ICRU 49
Lung 300 −726 0.29 0.287 0.287 0.287
Lung 450 −548 0.45 0.452 0.452 0.454
Adipose −95 0.94 0.955 0.960 0.963 0.946
Breast −46 0.99 0.990 0.993 1.001
Water −5 1.00 1.000 1.000 1.000
Brain 21 1.05 1.034 1.029 1.047 1.0287
Liver 1 1.10 1.074 1.077 1.078
Inner bone 223 1.15 1.112 1.140 1.123
B 200 bone material 242 1.16 1.121 1.149 1.130
CB 2%–30% CaCO3 471 1.34 1.262 1.262 1.266
CB 2%–30% CaCO3 845 1.56 1.431 1.447 1.427
Cortical bone 1263 1.82 1.626 1.650 1.620 1.659
a

Private communication with RaySearch Laboratories.

b

Stoichiometric cal. performed on‐site by an indepdent proton physicist.

c

Calculated using Eq. 13 of Fippel and Soukup.