Skip to main content
. 2017 Nov 10;19(1):73–85. doi: 10.1002/acm2.12222

Table 2.

Gamma passing rate comparisons of matched beam profiles

Energy Gamma Analysis (linac 1 vs linac 2) Gamma Analysis (linac 1 vs linac 3)
Crossline Inline Crossline Inline
Precorrection Postcorrection Precorrection Postcorrection Precorrection Postcorrection Precorrection Postcorrection
6 MV 89% 99% 53% 96% 95% 99% 56% 93%
10 MV 93% 99% 85% 95% 92% 92% 58% 93%
6 MeV 83% 98% 35% 86% 70% 96% 52% 82%
20 MeV 74% 95% 47% 96% 94% 96% 48% 92%