Table 1. The number of cycles to failure and length (mm) of fractured fragments of instruments during static cyclic fatigue testing in double curvature .

Apical Curvature Coronal Curvature
Group No. Number of Cycles to Fracture Fracture Length Number of Cycles to Fracture Fracture Length
WaveOne Gold 20 1028.12±164.48a 2.37±0.53a 1215.65±218.72a 6.35±1.22a
Reciproc Blue 20 987.45±148.05a 2.59±0.47a 1161.76±174.15a 6.48±1.07a
Genius File 20 4855.22±631.15b 2.42±0.49a 5730.11±916.85b 6.51±1.14a
P-value <0.05 >0.05 <0.05 >0.05