Table 1.

UK Biobank brain MRI protocol (31 min total scan time).

Modality Duration Voxel, Matrix Key Parameters #Volumes/#Timepoints
T1 4:54 1×1×1 mm 208×256×256 3D MPRAGE, sagittal, R = 2, TI/TR = 880/2000 ms 1
T2 FLAIR 5:52 1.05×1.0×1.0 mm 192×256×256 FLAIR, 3D SPACE, sagittal, R = 2, PF 7/8, fat sat, TI/TR = 1800/5000 ms, elliptical 1
swMRI 2:34 0.8×0.8×3.0 mm 256×288×48 3D GRE, axial, R = 2, PF 7/8 TE1/TE2/TR = 9.4/20/27 ms 2
dMRI 7:08 2.0×2.0×2.0 mm 104×104×72 MB = 3, R = 1, TE/TR = 92/3600 ms, PF 6/8, fat sat, b = 0 s/mm2 (5x + 3× phase-encoding reversed), b = 1 000 s/mm2 (50×), b = 2000 s/mm2 (50×) 105 + 6 (AP + PA)
rfMRI 6:10 2.4×2.4×2.4 mm 88×88×64 TE/TR = 39/735 ms, MB = 8, R = 1, flip angle 52°, fat sat 490
tfMRI 4:13 2.4×2.4×2.4 mm 88×88×64 Acquisition same as rfMRI. Task is faces/shapes “emotion” task. 332

R = in-plane acceleration factor, MB = multiband factor, PF = partial Fourier. All non-EPI scans are pre-scan normalised (on-scanner bias-field corrected). Gradient distortion correction is deselected on the scanner and applied in post-processing. SNR, tSNR and motion plots are shown in Figs. S10 and S11 of the supplementary material.