Skip to main content
. 2017 Dec 4;9(1):55–71. doi: 10.1364/BOE.9.000055

Table 2.

Results and uncertainties (k = 2) of B0430, t = 6.95 mm: μa, the absorption coefficient of the sample; μs the reduced scattering coefficient of the sample. The uncertainties on the INO results were estimated from measurements made on different samples at λ = 600 nm as presented in Ref. [3].

λ(nm) μa × 10−3 (mm−1) μs (mm−1)

INO NIST INO NIST
475 15.0 ± 1.7 14.6 ± 0.6 1.04 ± 0.07 1.09 ± 0.13
540 12.0 ± 1.5 12.5 ± 0.5 1.02 ± 0.07 1.09 ± 0.13
543 11.9 ± 1.3 12.5 ± 0.5 1.00 ± 0.07 1.08 ± 0.13
630 12.2 ± 1.4 11.5 ± 0.5 1.01 ± 0.07 1.06 ± 0.12
632 12.0 ± 1.4 11.5 ± 0.5 0.997 ± 0.070 1.05 ± 0.12
780 9.5 ± 1.1 9.5 ± 0.5 0.954 ± 0.067 1.00 ± 0.12
805 9.3 ± 1.1 9.2 ± 0.5 0.948 ± 0.066 0.98 ± 0.11
850 9.0 ± 1.0 8.9 ± 0.6 0.928 ± 0.065 0.97 ± 0.11