Table 3.
Results and uncertainties (k = 2) of the reduced scattering coefficient of the PDMS samples. The integrating sphere results are compared to time domain measurements by PTB.
λ (nm) | (mm−1) | ||
---|---|---|---|
| |||
PTB | NIST | ||
TiO2 = 0.05% | |||
| |||
500 | 0.590 ± 0.030 | 0.582 ± 0.052 | |
600 | 0.518 ± 0.030 | 0.517 ± 0.045 | |
700 | 0.444 ± 0.030 | 0.449 ± 0.039 | |
800 | 0.384 ± 0.030 | 0.391 ± 0.035 | |
TiO2 = 0.1% | |||
| |||
500 | 1.21 ± 0.06 | 1.19 ± 0.11 | |
600 | 1.04 ± 0.05 | 1.05 ± 0.10 | |
700 | 0.886 ± 0.044 | 0.904 ± 0.083 | |
800 | 0.773 ± 0.039 | 0.776 ± 0.071 | |
TiO2 = 0.2% | |||
| |||
500 | 2.30 ± 0.12 | 2.46 ± 0.24 | |
600 | 2.07 ± 0.10 | 2.17 ± 0.21 | |
700 | 1.77 ± 0.09 | 1.86 ± 0.18 | |
800 | 1.54 ± 0.08 | 1.57 ± 0.15 |