Table 2.

Areas under the receiver operating characteristic (AUC) curves and sensitivities at fixed specificities to discriminate glaucoma from healthy eyes for the nGoggle global multifocal steady-state visual evoked parameter and standard automated perimetry (SAP) parameters.

AUC (95% CI) Sensitivity for Specificity at 80% Sensitivity for Specificity at 90%
nGoggle global mfSSVEP 0.924 (0.863 – 0.964). 85% (71% – 95%) 71% (53% – 87%)
SAP MD 0.813 (0.716 – 0.896) 64% (45% – 82%) 43% (23% – 68%)
SAP MS 0.797 (0.687 – 0.880) 60% (39% – 77%) 39% (18% – 59%)
SAP PSD 0.768 (0.657 – 0.858) 61% (40% – 75%) 47% (24% – 66%)

Abbreviations: MS – mean sensitivity, MD – mean deviation; PSD – pattern standard deviation; mfSSVEP – multifocal steady-state visual evoked potential