Table I.

Representative accumulated data for laser eyewear test samples A – V with laser oscillator tuned to various center wavelengths, bandwidths, output power and impinging beam diameter. The maximum attainable system optical density (Max System OD) is also noted for each set of laser conditions.

Center Wavelength (nm) 847 850 792 789 800 800 800 800
Bandwidth (nm FWHM) 42.5 37 35.1 32.7 26.2 26.2 26.2 26.2
Average Power (mW) 238 229 128 125 470 400 420 375
Max System OD 6.42 6.41 6.40 6.41 8.88 9.02 8.82 9.07
Beam Diameter (mm) 2 10 10 10 15 15 15 2.5
Sample ODa d (mm)b OD Damc OD OD OD OD OD OD OD Damc
A >7 3.2 Infd Inf 8.88 9.07
B >7 3.2 Inf Inf 8.9 9.07
C 7+ 3.5 Inf Inf 7.21 7.78
D 7+ 7.0 Inf Inf 7.37 7.52 7.83
E 7+ 3.5 Inf Inf 8.88 9.07
F 2 3.2 4.83 3.28 3.27 4.32 3.5 Y
G 7+ 3.0 2.31 Y 5 7.62 7.38 Y
H 5+ 2.0 0.49 1 1.13 1.23
I 2.0 Inf Inf 8.89 8.18 Y
J >7 6.2 Inf Inf 8.9 8.15
K >6 3.4 Inf Inf 8.9 8.68 7.92
L >6 4.0 Inf Inf 8.9 8.88
M 7+ 3.0 2.35 Y 2.2 5.49 8.42 7.38 Y
N 5+ 3.0 1.93 Y 1.84 3.93 5.31 5.53 5.44
O 5+ 3.2 1.08 1.05 2.81 3.14 3.16
P 7+ 3.0 Inf Y Inf Inf 8.9 9.05 Y
Q 6+ 3.2 Inf Y Inf Inf 8.89 8.85 Y
R 7+ 3.0 Inf Y Inf 5.65 7.24 7.44 Y
S 4.3 6.4 4.77 4.73 3.04 5.11 Y
T >7 2.0 2.5 5.13 7.91 5.91 Y
U 7+ 2.0 7.46 8.11 7.33 Y
V 2.0 7.53 7.86 7.47 Y
a

Optical Density at 800 nm from Manufacturer either with sample or extracted from website

b

No filter thickness provided on supplier’s websites so nominal measured thickness (d) supplied

c

“Y” in Dam columns signifies damage to sample was observed (discoloration or distortion) during the course of several seconds of exposure under the experimental conditions specified in the preceding column

d

“Inf” = infinity; signifies measured OD exceeds the system dynamic range maximum value