Published in final edited form as: Sci Total Environ. 2017 Nov 6;618:579–585. doi: 10.1016/j.scitotenv.2016.12.123

Table 1.

Summary of statistical results for both laboratory and field analyses. Slope, intercept, and r2 coefficient were calculated from linear regression model of photo predicted versus ICP-MS determined As for concentrations up to 150 μg/L, and for full range with maximum As concentration display to the right. Laboratory and field detection limits (DL) based on predicted values from regression up to 150 μg/L, and full range. Field blanks were classified by ICP-MS determined concentrations below 1 μg/L. If there were less than 5 blanks, the detection limit was not provided.

0–150 μg/L Full range
Date Labe1 Figure n r2 DL
(μg/L)
n r2 DL
(μg/L)
Max As
(ug/L)
April 16th, 2015 Lab 1 SI-1 5 1.00 9 0.96 1000
April 23rd, 2015 Lab 2 SI-2 13 0.96 17 0.99 1000
Oct, 7th, 2015 Lab 3 SI-3 14 0.97 24 0.95 1280
Jul 30th, 2016 Lab 4 SI-4 10 0.99 8.2 13 0.97 150 1000
2015–2016 Lab 1–4 SI-5 42 0.96 9.2 62 0.90 180 1280
July, 2015 Field SI-6 220 0.83 22 274 0.75 61 750