Table 2.

Interobserver reproducibility outcomes for biometric measurements obtained using AL-Scan partial coherence interferometry in children.

Parameter Mean difference ± SD Sw TRT CoV (%) ICC
AL (mm) 0.01 ± 0.03 0.02 0.06 0.09 1.000
CCT (μm) −0.02 ± 1.51 1.06 2.93 0.19 0.999
ACD (mm) 0.00 ± 0.01 0.01 0.02 0.22 0.998
Kf (D) −0.02 ± 0.09 0.08 0.21 0.17 0.998
Ks (D) −0.02 ± 0.11 0.06 0.18 0.15 0.998
Km (D) −0.02 ± 0.08 0.06 0.16 0.13 0.998
Astigmatism 0.01 ± 0.12 0.09 0.24 6.68 0.989
J0 (D) 0.00 ± 0.06 0.04 0.11 0.990
J45 (D) 0.00 ± 0.05 0.03 0.09 0.976
CD (mm) 0.03 ± 0.10 0.07 0.21 0.62 0.955
PD −0.01 ± 0.11 0.08 0.22 1.01 0.981

AL = axial length, CCT = central corneal thickness, ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, CD = corneal diameter, SD = standard deviation, Sw = within-subject standard deviation, TRT = test-retest repeatability (2.77 Sw), CoV = within-subject coefficient of variation, ICC = intraclass correlation coefficient.