Skip to main content
. 2018 Jan 22;8:1304. doi: 10.1038/s41598-018-19605-6

Table 3.

The mean difference, paired T-test, and 95% limits of agreement (LoA) for differences between the AL-Scan and the IOLMaster partial coherence interferometry in children.

Device Pairings Mean Difference ± SD P Value 95% LoA
AL (mm) 0.00 ± 0.03 0.346 −0.05 to 0.05
ACD (mm) 0.17 ± 0.06 0.000 0.05 to 0.29
Kf (D) −0.08 ± 0.14 0.000 −0.36 to 0.20
Ks (D) −0.19 ± 0.17 0.000 −0.52 to 0.14
Km (D) −0.13 ± 0.12 0.000 −0.37 to 0.10
Astigmatism −0.11 ± 0.20 0.000 −0.50 to 0.27
J0 (D) 0.02 ± 0.19 0.537 −0.37 to 0.40
J45 (D) −0.01 ± 0.12 0.608 −0.25 to 0.24
CD (mm) −0.42 ± 0.35 0.000 −1.10 to 0.26

AL = axial length, ACD = anterior chamber depth, Kf = flattest keratometry, Ks = steepest keratometry, Km = mean keratometry, CD = corneal diameter, SD = standard deviation.