Fig. 5.

Fig. 5

Typical trial in experimental group 2: non-irrigated electrode, blood flow 0.26 m/sec. Note that both T0 and TE rise rapidly to boiling, followed rapidly by a marked impedance rise and automatic generator shutoff.