Skip to main content
. 2018 Jan 24;56(2):e01237-17. doi: 10.1128/JCM.01237-17

TABLE 2.

Xpert Flu+RSV Xpress Assay performance versus ProFlu+ on NP swab specimens for CW sites and non-CW sites combined

Specimen type (n)a Target No. of results by typeb
PPA (% [95% CI]) NPA (% [95% CI]) Accuracy (% [95% CI])
TP FP TN FN
Fresh (2,176) Influenza A virus 250 101 1,825 0 100 (98.5–100) 94.8 (93.7–95.7) 95.4 (94.4–96.2)
Influenza B virus 63 10 2,103 0 100 (94.3–100) 99.5 (99.1–99.8) 99.5 (99.2–99.8)
RSV 125 8 2,039 4 96.9 (92.3–99.1) 99.6 (99.2–99.8) 99.4 (99.0–99.7)
Preselected frozen (259) Influenza A virus 0 4 255 0 NAc 98.5 (96.1–99.6) 98.5 (96.1–99.6)
Influenza B virus 100 2 157 0 100 (96.4–100) 98.7 (95.5–99.8) 99.2 (97.2–99.9)
RSV 40 1 217 1 97.6 (87.1–99.9) 99.5 (97.5–100) 99.2 (97.2–99.9)
Fresh and preselected frozen combined (2,435) Influenza A virus 250 105 2,080 0 100 (98.5–100) 95.2 (94.2–96.1) 95.7 (94.8–96.5)
Influenza B virus 163 12 2,260 0 100 (97.8–100) 99.5 (99.1–99.7) 99.5 (99.1–99.7)
RSV 165 9 2,256 5 97.1 (93.3–99.0) 99.6 (99.2–99.8) 99.4 (99.0–99.7)
a

Fresh, specimen prospectively collected from consented/assented subjects; preselected frozen, specimen deidentified, frozen, and known to be positive or negative for a specific analyte. n, number of specimens.

b

TP, true positive; FP, false positive; TN, true negative; FN, false negative.

c

NA, not applicable.