Table 2.
Agreement of MD and PSD between Matrix Frequency-Doubling Perimetry (FDT) and Swedish Interactive Thresholding Algorithm (SITA) Standard Automated Perimetry (SAP)
| Parameter of interest | Correlation of two sets | ∆ Mean ± SD | ∆ Range | Absolute ∆ Mean ± SD | P-value * | 95% CI | ICC | LoA (lower to upper) |
| MD | 0.259 | 0.1 ± 3.4 | -8.1 to 9.9 | 2.5 ± 2.2 | 0.905 | -0.94 to 1.06 | 0.06 | -6.56 to 6.76 |
| PSD | 0.071 | -0.5 ± 2.5 | -5 to 8.3 | 1.9 ± 1.6 | 0.169 | -1.24 to 0.22 | 0.24 | -5.4 to 4.4 |
: Difference between SITA-SAP and Matrix-FDT, LoA: limit of agreement, MD: mean deviation, PSD: pattern standard deviation, SD: standard deviation, CI: confidence interval,
Based on the paired t-test