Skip to main content
. 2017 Nov 29;9(11):e1889. doi: 10.7759/cureus.1889

Table 2. Predictive errors for the attenuation model created with values normalized to 10 cm of attenuation.

  Predictive errors for attenuation (cm)
MU/Beam 0 5 10 16 22
10 0.4% (0.68%) 1.4% (1.19%) 0.1% (1.45%) 0.6% (1.98%) -0.3% (2.52%)
20 -0.1% (0.27%) 0.0% (0.51%) 0.0% (0.69%) 0.4% (1.70%) -2.6% (1.28%)
50 -0.1% (0.15%) 0.3% (0.37%) 0.0% (0.40%) 0.6% (0.60%) -0.4% (0.88%)
100 -0.2% (0.14%) 0.3% (0.20%) -0.1% (0.37%) 0.6% (0.31%) -0.4% (0.60%)
200 -0.3% (0.14%) 0.1% (0.19%) 0.0% (0.19%) 0.9% (0.51%) -0.8% (0.67%)