Table 1.

A summary of recommended settings for robust optimization of craniospinal irradiation.

Action Recommendation
OTV Expansion from CTV 0 mm (OTV = CTV) in the brain region; 10 mm sup/inf in the spine to eliminate scalloping; 2 to 4 mm right/left and ant/post in the spine region.
STV Expansion from fs_OTV Desired robustness + spot spacing (i.e., 3 mm + 5 mm), 3% range
Plan Perturbations +/− 3%, +/− 3 mm in each direction
Field Perturbations 5 mm sup/inf for each field in the match regions.
Robustly optimized structures OTV, cribriform brain, optic structures, brainstem, cochlea (optional)

fs OTV: field specific Optimization Target Volume; CTV: Clinical Target Volume; STV: Scanning Target Volume. +/−: plus/minus.