Table 2.

Intra-day accuracy and precision data of the analytes.

Qualitycontrol Run PIO
TLM
Mean found (ng/mL) SD % CV Accuracy % Mean found (ng/mL) SD % CV Accuracy %
LLOQ 1 5.15 0.19 3.71 102.92 4.90 0.50 10.14 97.92
2 5.20 0.23 4.33 103.96 5.09 0.46 9.13 101.76
3 5.27 0.17 3.19 105.44 4.86 0.40 8.21 97.24
4 4.79 0.32 6.74 95.84 5.18 0.27 5.30 103.68



LQC 1 15.24 0.93 6.11 101.60 14.46 0.83 5.72 96.40
2 14.32 1.00 7.01 95.47 14.46 1.07 7.39 96.40
3 14.52 1.04 7.15 96.80 14.14 0.98 6.92 94.27
4 14.72 0.53 3.58 98.13 14.86 0.57 3.82 99.07



MQC 1 4404.00 168.02 3.82 97.87 4538.00 302.27 6.66 100.84
2 4644.00 254.13 5.47 103.20 4734.00 309.24 6.53 105.20
3 4302.00 149.23 3.47 95.60 4616.00 275.92 5.98 102.58
4 4388.00 199.67 4.55 97.51 4630.00 107.47 2.32 102.89



HQC 1 8072.00 687.95 8.52 100.90 7650.00 491.07 6.42 95.63
2 7782.00 654.46 8.41 97.28 7918.00 442.12 5.58 98.98
3 8488.00 422.46 4.98 106.10 8076.00 706.14 8.74 100.95
4 7742.00 455.16 5.88 96.78 7868.00 463.86 5.90 98.35

% CV: coefficient of variation. Accuracy: (Mean assayed concentration – nominal concentration)/(nominal concentration) × 100.