Skip to main content
. 2017 Dec 30;8(1):17. doi: 10.3390/nano8010017

Table 1.

Comparison of limits of detection (LOD) and sensitivity values with respect to DA, AA, and UA as determined by amperometric and DPV analysis.

Targets DA AA UA
Amperometry Linear regression equation ip (μA) = 7.6 × 10−8 + 0.43C (μM)
R2 = 0.954
(0.14–100 μM),
ip (μA) = 1.4 × 10−8 + 0.023C (μM)
R2 = 0.968
(100–700 μM)
ip (μA) = 4.8 × 10−8 + 0.0081 C (μM)
R2 = 0.988
ip (μA) = 2.1 × 10−8 + 0.0035 C (μM)
R2 = 0.979
LOD (μM) 0.14 9.5 25
Sensitivity (μA·μM−1) 0.43 8.1 × 10−3 3.5 × 10−3
DPV Linear regression equation ip (μA) = 1.8 × 10−7 + 0.27 C (μM)
R2 = 0.935
ip (μA) = 1.2 × 10−7 + 0.0031 C (μM)
R2 = 0.989
ip (μA) = 6.0 × 10−7 + 0.051 C (μM)
R2 = 0.977
LOD (μM) 0.69 5.7 2.2
Sensitivity (μA·μM−1) 0.27 3.1 × 10−3 0.051