Published in final edited form as: IEEE Trans Biomed Eng. 2017 Apr 12;65(1):52–63. doi: 10.1109/TBME.2017.2688919

Table II.

Estimation Biases (%) of Sensor Parameters with the Seperate Calibration Algorithm (S) [9] and with the JCL (J). The Lowest Bias for Each Dataset and Parameter Is in Bold Italic.

Evaluation Gain
Euler Angle
Offset
Datasets S J S J S J
D2 4.37 3.50 2.62 2.72 328.53 69.66
D3 10.18 7.90 2.55 2.40 28.87 21.69
D4 8.32 6.81 2.90 2.80 67.04 30.63
D5 5.54 2.38 1.87 1.48 40.23 15.48
D6 6.20 6.40 4.09 4.03 128.20 34.12
D7 13.99 9.24 4.17 4.01 189.03 83.21
D8 7.98 5.74 3.56 3.63 147.61 35.41
D9 10.64 6.31 3.56 3.44 184.63 37.50
D10 13.65 13.20 4.60 4.43 289.97 113.99
D11 8.48 8.60 4.60 4.67 136.00 44.94
D12 14.72 11.18 4.39 4.34 171.58 99.19
D13 10.00 8.03 4.02 4.09 166.67 54.98
D14 11.69 6.15 4.49 4.30 197.67 63.12
D15 16.08 15.91 5.84 5.28 226.83 133.85

Avg. 10.13 7.95 3.81 3.69 164.49 59.84
Std. 3.58 3.60 1.04 1.02 85.24 36.05