Skip to main content
. 2017 Nov 16;2:22. doi: 10.1186/s41077-017-0057-y

Table 2.

Effect of source of feedback on effective compression depth

Effective compression depth (mm) Mean (95% confidence interval)
Source of feedback
Anterior sensor Internal measure Difference
Typical mattress only 37.8 (34.1, 41.5) 52.1 (50.1, 54.1) 14.3 (12.0, 16.5)
Typical mattress + backboard 42.9 (39.5, 46.3) 51.7 (49.8, 53.5) 8.7 (6.5, 11.0)
Memory foam mattress only 41.2 (37.9, 44.4) 54.2 (52.3, 56.1) 13.0 (10.7, 15.3)
Memory foam mattress + backboard 46.3 (43.2, 49.3) 53.7 (52.0, 55.5) 7.5 (5.2, 9.8)

Presented are the mean adjusted values from a mixed effects linear model. Fixed effect: mattress type, CPR backboard use, feedback source, interaction between feedback source and mattress type, and interaction between feedback source and CPR backboard use. Random effect: intercept, mattress type, CPR backboard use, and feedback source. Goodness of fit: marginal R 2 = 0.43, conventional R 2 = 0.80