Skip to main content
. 2017 Dec 21;45(2):884–897. doi: 10.1002/mp.12699

Table 1.

Deviations from the beam central axis (CAX) of the center of the dose profiles measured by various detectors in a 1.5 T magnetic field in both the inline direction (parallel with the magnetic field) and the crossline direction (perpendicular to the magnetic field). CAX deviations in the crossline direction are shown based on the value measured from the raw data and the value when corrected for the effects of volume averaging. Uncertainty is ±0.1 mm. Positive values indicate a shift toward the average direction of the Lorentz force (patient left when head‐first supline)

Detectors CAX deviations (mm)
Inline Crossline (raw value) Crossline (corrected)
PTW 30013 Farmer N/A 2.36 2.16
PTW Semiflex 3D −0.07 2.05 1.92
PTW PinPoint 3D 0.04 1.07 1.01
PTW microDiamond 0.09 0.83 0.79
PTW Diode SRS 0.01 0.83 0.82
PTW Diode P 0.01 0.55 0.54