Table 2: Strengths and Weaknesses of Embolic Protection Devices.

Proximal Occlusion Distal Occlusion Distal Filter
Maintenance of antegrade blood flow during intervention - - +
Limited contrast opacification + + -
Unlimited debris capture + + -
Capture of debris <100 μm + + -
Capture of soluble mediators + + -
Shunting of debris into proximal side branches - + -
Ease of use Complex Complex Simple
Manoeuverability Good Good Reduced
Crossing profile NA Low (2.7 Fr)* High (3.2 Fr)**

NA = not available.

*PercuSurge GuardWire (Medtronic).

**FilterWire EZ (Boston Scientific).