Published in final edited form as: Magn Reson Med. 2017 Aug 30;79(4):2077–2086. doi: 10.1002/mrm.26881

Table 1.

Imaging Parameters used for Ungated Nonenhanced QISS FISS MRA

Parameter Pelvic MRA Carotid MRA Intracranial MRA
FISS TRf (ms) 17.6 17.3 10.7
TRb (ms) 3.8 3.9 4.0
TE (ms) 1.9 1.9 2.0
Flip angle (°) 70 70 67
FISS n 3 3 1
FISS RF spoiling yes no no
QISS TI (ms) 1600 505 400
Field of view (mm × mm) 400 × 400 256 × 256 256 × 256
Reconstructed Matrix 384 × 384 256 × 256 256 × 256
rn, radial views 147/441a 132 192
rθ, view angle increment (°) 15.918/15.102a 17.727 15.938
Radial undersampling factorb 4.1/1.4a 3.0 2.1
Slice thickness (mm) 3 3 1
Slice overlap (mm) 1.5 0.5
Imaging shots per slice 1/3a 1 4
Bandwidth (Hz/pixel) 930 930 975
Number of slices 120 120 120
Scan time (per slice) 2.07/6.21 sa 1.01 s 2.58 s
Scan time (total) 9 min 40 s 2 min 1 s 5 min 10 s
FISS Gspoil duration (ms) 1.4 5.1
FISS Gspoil strength (mT/m) 7.8 7.8
a

Values provided in the form: lower 40 slices / upper 80 slices.

b

With respect to satisfying the Nyquist sampling requirements. FISS Gspoil refers to an additional gradient applied between FISS modules (slice axis). n=number of bSSFP views acquired per FISS module; QISS=quiescent-interval slice-selective; TI=inflow time. Field of view and matrix values are after removal of oversampling in readout direction and without zero-filling interpolation.