Skip to main content
. 2018 Feb 13;9:655. doi: 10.1038/s41467-017-02714-7

Table 2.

HPLC conditions for analysis of digested and undigested synthetic RNA oligonucleotides

Column: Waters XSelect HSS T3, 100 Å, 3.5 µm, 3.0 × 150 mm
Column temperature: 30 °C
Flow rate: 0.7 mL min−1
Injection volume: 8 μL
Step time / min Mobile phase composition / %
A (water) B (acetonitrile) C (0.2 M NH4OAc, pH 6, 25% acetonitrile)
0 89 1 10
2.5 89 1 10
5.5 80 10 10
10 40 50 10
10.1 10 80 10
11 10 80 10
11.1 89 1 10
20 89 1 10