| OCT platform |
AngioVue RTVue XR Avanti |
CIRRUS HD-OCT Model 5000 |
Spectralis OCT-2 |
DRI-OCT Triton swept source OCT |
RS-3000 Advance |
| Imaging company |
Optovue |
Carl Zeiss Meditec, Inc |
Heidelberg Engineering |
Topcon Corporation |
Nidek |
| Place of origin |
Fremont, CA, USA |
Dublin, CA, USA |
Heidelberg, Germany |
Tokyo, Japan |
Gamagori, Aichi, Japan |
| Scanning speed |
70 000 scans/s |
68 000 scans/s |
85 000 scans/s |
100 000 scans/s |
53 000 scans/s |
| Scanning volume |
304 × 304 A scans |
245 × 245, 350 × 350
A scans |
512 × 512 A scans |
320 × 320, 512 × 512 A scans |
256 × 256 A scans |
| Algorithm |
SSADA |
OMAG |
Probalistic model that predicts whether a voxel contained flow or not. |
OCTA—ratio analysis (full-spectrum amplitude) |
Complex difference (full-spectrum amplitude) |
| Type of algorithm |
Amplitude |
Amplitude+phase |
Probablistic model |
Amplitude |
Amplitude +phase |
| Scan area (macula) |
3 × 3, 6 × 6, 8 × 8 mm |
3 × 3, 6 × 6, 8 × 8 mm |
3 × 3 mm with (5.7 × 5.7) μm/px |
3 × 3, 4.5 × 4.5, 6 × 6, 9 × 9 mm |
3 × 3–9 × 9 mm (12 × 9 montage) |
| |
|
|
|
|
|
| Optical resolution |
| Axial |
3 μm |
5 μm |
7 μm |
8 μm |
7 μm |
| Lateral |
15 μm |
15 μm |
14 μm |
20 μm |
20 μm |
| Light source |
840 nm |
840 nm |
880 nm |
1050 nm |
880 nm |
| Axial imaging depth |
2–3 mm |
2 mm |
1.9 mm |
2.6 mm |
2.1 mm |
| Automated segmentation options |
Superficial retinal capillary plexus
Deep retinal capillary plexus
Outer retina
Choriocapiliaries |
Retina depth encoded
Vitreo-retinal interface
Superficial retina
Deep retina
Avascular layer
Choriocapillaris
Choroid |
Four presets matching vasculature in retinal nerve fibre layer, ganglion cell layer and bracketing the inner nuclear layer
Three presets to cover the retina (superficial, deep vascular plexus and avascular layer) |
|
|
| |
Superficial vascular plexus
Deep vascular plexus
Outer retina
Choriocapillaris |
Superficial retinal layer
Deep retinal layer
Avascular Choriocapillaris |
|
|
|
| Colour coding of segmentations |
Yes |
Yes |
Yes |
Yes |
Yes |
| Cross-sectional OCTA image |
Yes |
Yes |
Yes |
Yes |
No |
| Eye tracker |
Software update for older models
Available with newer models |
Yes (Fast Trac) |
Yes (TruTrack) |
Yes (Smart Track) |
Yes |
| Motion correction |
Yes (motion tracker) |
N/A |
Covered by TruTrack |
Yes |
Yes (real-time SLO eye tracking) |
| Projection artefact removal |
Yes |
Yes |
Under development |
Yes |
Yes |
| Optic nerve OCTA |
Yes |
Yes |
Yes |
Yes |
Yes |
| Anterior segment OCTA function |
Prototype |
Under development |
No |
No |
Yes |
| Quantitative analysis |
Yes |
Yes |
Under development |
Yes (prototype) |
Yes |
| Comparative follow-up function |
Yes |
Yes |
Yes |
Yes |
No |