Skip to main content
. 2017 Sep 13;118(6):3271–3281. doi: 10.1152/jn.00436.2017

Table 1.

Array specifications for each animal

Monkey ID Array Type Size, mm Electrode Length, mm Manufacturing Impedance, MΩ Array Signal Longevity Likely Failure Mode
WH FMA HD 1.6 × 2.95 2.0–2.5 0.5–0.9 6 wk Wire bundle failure
TE FMA HD 1.6 × 2.95 2.0–2.5 0.3–1.1 6 mo Wire bundle failure
PI FMA SD 1.8 × 4 1.5–2.0 0.4–0.7 6 wk Wire bundle failure
BA FMA SD 1.8 × 4 2.0–2.5 0.7–1.1 3 wk Wire bundle failure
CU FMA HD 1.6 × 2.95 2.0–2.5 0.4–1.0 N/A Fluid leakage
CH2 FMA SD 1.6 × 2.8 1.5–3.0 0.08–1.5 <1 wk Array manufacturing defect
KR FMA SD 1.8 × 4.0 1.0–2.0 0.3–1 2 wk Array came out
MR FMA HD 1.6 × 2.8 1.25–2.5 0.6–1.3 N/A Array did not insert
HA FMA SD 1.8 × 4.0 1.0–2.25 0.5–1.0 6 wk Wire bundle failure
LA1 FMA HD 1.6 × 2.8 1.0–2.5 0.5–1.0 2 mo Wire bundle failure
OL Utah 4.0 × 4.0 1.5 0.3–0.8 1 day Meningoencephalitis
LA2 Utah 4.0 × 4.0 1.5 0.1–0.8 >6 mo N/A
CH1 Utah 4.0 × 4.0 1.3 0.2–0.8 <1 wk Array came out

HD, high definition; SD, standard definition; N/A, not applicable.