. 2018 Feb 1;2018:6148515. doi: 10.1155/2018/6148515

Table 2.

Validation data: calibration curve equation, correlation coefficient, LLOQ, ULOQ, intra- and interday precision, and average concentration ± SD quality control samples (human plasma from 6 volunteers) for L-arginine and ADMA using the selected sample preparation (A) with the developed chromatographic method.

Calibration curve equation Correlation coefficient (R2) LLOQ (µM) ULOQ (µM) Intraday precision (CV, %) (n = 6) Interday precision (CV, %) (n = 9) Quality control (µM) (n = 6)
L-arginine y = 78686x − 3519 0.999 0.14 15 3.2 5.0 64.1 ± 10.3
ADMA y = 73655x + 1134 0.994 0.012 0.50 3.4 3.3 0.27 ± 0.02
SDMA y = 87003x + 3261 0.989 4.2 3.8 0.57 ± 0.09