Table 3.
Device Parameter Changes Immediately after Cardiac Magnetic Resonance Imaging Study and at Follow-up Compared to Pre-scan Values
Parameter | Post change | Post percent change | p-value | Follow-Up change | Follow-up percent change | p-value |
---|---|---|---|---|---|---|
P-wave amplitude (mV) | 0 (−0.3 to 0.1) | 0 (−11 to 5) | 0.256 | 0 (−0.4 to 0.7) | 2.7 (−11 to 37) | 0.182 |
R-wave amplitude (mV) | 0 (−0.4 to 0.3) | 0 (−5 to 3) | 0.492 | 0 (−0.6 to 0.4) | 0 (−8 to 5) | 0.331 |
Atrial-lead impedance (Ω) | −7 (−20 to 0) | −2 (−4 to 0) | <0.0001 | −18.5 (−41 to 6) | −4 (−8 to 1) | 0.007 |
Right-ventricular lead impedance (Ω) | 0 (−19 to 0) | 0 (−4 to 0) | 0.0001 | −19 (−44 to 7) | −2.7 (−9 to 2) | 0.006 |
Left-ventricular lead impedance (Ω) | −10 (−30 to 0) | −3 (−6 to 0) | 0.023 | −30 (−130 to 0) | −8 (−18 to 0) | 0.003 |
Shock lead impedance (Ω) | 0 (−2.5 to 1) | 0 (−3 to 2) | 0.100 | −2 (−4 to 1.3) | −3 (−7 to 2) | 0.219 |
Atrial capture threshold (V)† | 0 (0 to 0) | 0 (0 to 0) | 0.847 | 0 (−0.1 to 0.125) | 0 (−12.5 to 20) | 0.440 |
Right-ventricular capture threshold (V)† | 0 (0 to 0) | 0 (0 to 0) | 0.072 | 0 (0 to 0.175) | 0 (0 to 17) | 0.100 |
Left ventricular capture threshold (V)† | 0 (0 to 0.175) | 0 (−18 to 6) | 0.4875 | −0.05 (−0.22 to 0.10) | −3 (−10 to 13) | 0.336 |
Battery‡ | ||||||
Voltage (V) | 0 (0 to 0) | 0 (0 to 0) | 0.1147 | |||
Longevity (years) | 0 (0 to 0) | 0 (0 to 0) | 0.116 |
Direct comparisons not performed if pulse width were different
Battery parameters were not evaluated at follow-up due to many other unrelated intervening factors which may affect this parameter