Published in final edited form as: Heart Rhythm. 2017 Oct 7;15(2):218–225. doi: 10.1016/j.hrthm.2017.10.003

Table 3.

Device Parameter Changes Immediately after Cardiac Magnetic Resonance Imaging Study and at Follow-up Compared to Pre-scan Values

Parameter Post change Post percent change p-value Follow-Up change Follow-up percent change p-value
P-wave amplitude (mV) 0 (−0.3 to 0.1) 0 (−11 to 5) 0.256 0 (−0.4 to 0.7) 2.7 (−11 to 37) 0.182
R-wave amplitude (mV) 0 (−0.4 to 0.3) 0 (−5 to 3) 0.492 0 (−0.6 to 0.4) 0 (−8 to 5) 0.331
Atrial-lead impedance (Ω) −7 (−20 to 0) −2 (−4 to 0) <0.0001 −18.5 (−41 to 6) −4 (−8 to 1) 0.007
Right-ventricular lead impedance (Ω) 0 (−19 to 0) 0 (−4 to 0) 0.0001 −19 (−44 to 7) −2.7 (−9 to 2) 0.006
Left-ventricular lead impedance (Ω) −10 (−30 to 0) −3 (−6 to 0) 0.023 −30 (−130 to 0) −8 (−18 to 0) 0.003
Shock lead impedance (Ω) 0 (−2.5 to 1) 0 (−3 to 2) 0.100 −2 (−4 to 1.3) −3 (−7 to 2) 0.219
Atrial capture threshold (V) 0 (0 to 0) 0 (0 to 0) 0.847 0 (−0.1 to 0.125) 0 (−12.5 to 20) 0.440
Right-ventricular capture threshold (V) 0 (0 to 0) 0 (0 to 0) 0.072 0 (0 to 0.175) 0 (0 to 17) 0.100
Left ventricular capture threshold (V) 0 (0 to 0.175) 0 (−18 to 6) 0.4875 −0.05 (−0.22 to 0.10) −3 (−10 to 13) 0.336
Battery
Voltage (V) 0 (0 to 0) 0 (0 to 0) 0.1147
Longevity (years) 0 (0 to 0) 0 (0 to 0) 0.116

Direct comparisons not performed if pulse width were different

Battery parameters were not evaluated at follow-up due to many other unrelated intervening factors which may affect this parameter