Skip to main content
. 2018 Feb 22;56(3):e01691-17. doi: 10.1128/JCM.01691-17

TABLE 1.

Sensitivities and specificities of Aries, Xpress, and Cobas assays

Virus Assay Test result/reference standard result
Sensitivity (%) (95%CI) Specificity (%) (95%CI) Kappa (95%CI)
+/+ +/− −/+ −/−
FluA Aries 48 1 2 143 96.0 (86.5–98.9) 99.3 (96.2–99.9) 0.96 (0.91–1.0)
Xpress 50 0 0 144 100.0 (92.9–100) 100.0 (97.4–100) 1.0 (1.0–1.0)
Cobas 50 0 0 144 100.0 (92.9–100) 100.0 (97.4–100) 1.0 (1.0–1.0)
FluB Aries 44 1 1 148 97.8 (88.4–99.6) 99.3 (96.3–99.9) 0.97 (0.93–1.0)
Xpress 44 0 1 149 97.8 (88.4–99.6) 100.0 (97.5–100) 0.99 (0.96–1.0)
Cobas 45 1 0 148 100.0 (92.1–100) 99.3 (96.3–99.9) 0.99 (0.96–1.0)
RSV Aries 48 0 1 145 98.0 (89.3–99.6) 100.0 (97.4–100) 0.99 (0.96–1.0)
Xpress 49 0 0 145 100.0 (92.7–100) 100.0 (97.4–100) 1.0 (1.0–1.0)
Cobas 49 1 0 144 100.0 (92.7–100) 99.3 (96.2–99.9) 0.99 (0.96–1.0)