Skip to main content
. 2018 Feb 23;12(1):011503. doi: 10.1063/1.5010158

TABLE I.

DEP-based cancer cell manipulation platforms and characteristics.

Target cancer cell-line (References) DEP platform characteristic DEP platform type Optimum operating conditions Optimum performance of the platform for target cells
Breast cancer cells
MCF-7 (Kim et al.97) 3D-asymmetric microelectrode dielectrophoretic activated cell sorter eDEP • Flow rate: 0.35 μl/min
• Frequency: 45 MHz • 84% cell recovery
• Voltage: 7 Vpp
MCF-7 (Bhattacharya et al.98) DC selective trapping iDEP platform iDEP • Electric field: 100 V/cm • 68.5% cell viability after 2 min exposure to the electric field
GX treated MCF-7-LCC9 (Soltanian-Zadeh et al.99) Off-chip passivated-electrode insulator based Dielectrophoresis for separation iDEP • Flowrate: 20 μl/h
• Drug dosage: 500 nM • GX-treated cells • 72% ± 1.1% trapping efficiency
• Frequency: 1 MHz
• Voltage: 80 Vpp
GFP Labelled MDA-MB-231 (Alazzam et al.86) Planar interdigitated transducer electrodes for separation eDEP • Flow rate: 6 μl/h
• Voltage: 15 Vpp • Separation accuracy: 100%
• Frequency: 40 kHz • Purity: around 81%
MDA-MB-231 (Li and Anand101) Wireless electrode array eDEP • Flow rate: 0.1 ml/h
• Voltage: 248 Vpp • N/A
• Frequency: 40 kHz
MDA-MB-231 (Henslee et al.16) cDEP fluid electrode channel cell trapping device cDEP • Flow rate: 0.02 ml/h
• Voltage: 20 Vrms • 90% or greater target cell trapping
• Frequency: 180–210 kHz
Leukemia
HL-60 (Su et al.87) DEP spring method using angled coplanar electrodes for electrical cell characterization eDEP σm: Min. - 1.47 S/m, max. - 0.35 S/m • Characterization of approximately 10 cells/s
• Algorithm accuracy of 96%
K562 (Lee et al.106) Gravitational and nDEP based electrode array cell sorter eDEP • Throughput: 17 000 cells/min
• Voltage: 7 Vpp • Recovery rate: 49.42%
• Frequency: 10 kHz • Sorting efficiency: 94.74 ± 0.77%
THP-1 (Yasukawa et al.107) nDEP based electrode with the navigator and separator configuration for cell sorting by size eDEP • Flow velocity: 0.2 mm/s
• Voltage: Navigator - 15 Vpp • Separator – 14 Vpp • 88% passing accuracy
• Frequency: 100 kHz
σm: 110 mS/m
Jurkat T-lymphoblasts (Novickij et al.108) High pulsed magnetic field inducing pDEP electric field coil cDEP • Frequency altered magnetic field: 1.3 MHz
• Average cell velocity: 73 μm/s • N/A
• Pulse repetitive frequency: 1–25 Hz
Prostate cells
ALDH expressing PC-3 cells (Salmanzadeh et al.122) pDEP based selective trapping device cDEP • Flow rate: 0.2 ml/h
• Voltage: 293 Vrms • Around 69% separation of PC-3 cells from non-PC-3 cells
• Frequency: 200 kHz
PC-3 live cells (Sun et al.114) Ionic liquid electrode device for target cell separation cDEP • Flow rate: 50 μl/h
• Voltage: 53 Vrms • 89.8% live cancer cells deflected to the designated outlet.
Frequency: 50 kHz
Live and dead PC-3 cells and blood (Chiu et al.20) nDEP optically induced dielectrophoresis cell manipulation with laminar flow CTC cell isolation ODEP • Flow rate: 2.5 μL/min
• Voltage: 8 V
• Frequency: 50 kHz • Separation purity: 100%
• Hollow circular light images: • Cancer cell recovery rate: 41.5%
• Fixed ID – 40 μm
• Bandwidth – 40 μm
• Moving velocity – 50 μm/s
22Rv1 cells (Khamenehfar et al.115) Crossflow microfilter-DEP integrated single cancer cell isolation chip eDEP • Voltage: 11.5 V
• Frequency: 3 MHz • Isolation time of cancer cells >60 s
Ovarian cells
MOSE cells (Čemažar et al.118) Cell-scale micropillar (20 μm) pDEP cell isolation and trapping device cDEP • Flow rate: 20 μl/min
• Voltage: 300 Vrms • 71%–90% cancer cell viability
• Frequency: 50 kHz • Sorting over 1 × 106 cancer cells/h
Cervical cells
HeLa cells from RBC (Chen et al.18) pDEP stepping electric field cancer cell concentration and collection open-top chamber device eDEP • Voltage: 8 Vpp
• Frequency: 1 MHz • 76%–80% cancer cell survival rate
• Relay switching interval: 20 s • 76%-80% cancer cell concentration
• Cell concentration: 500 cells/ml
HeLa cells (Das et al.119) 3-inlet 3-outlet microfluidic platform with a tapered planar microelectrode eDEP • Voltage: 10 Vpp
• Frequency: 1 MHz • N/A
• Cell concentration: 106 cells/ml
Colorectal cells
HCT116 cells (Xing and Yobas120) Continuous flow IDT-3D ring array microelectrode (40 μm) eDEP • Flow rate: 0.2 ml/h
• Voltage: 10 Vpp • 94% cancer cell viability
• Frequency: 100 kHz • 82% cancer cell recovery
• Cell concentration: 107 cells/ml
Lung cells
AS2-GFP from RBC and WBC cells (Cheng et al.88) 3D LDEP nDEP lateral displacement CTC isolation system eDEP • Flow rate: 10 μl/min
• Voltage: 18 Vpp • Recovery was greater than 85% and the isolation
• Frequency: 10 kHz • Purity was greater than 90%
• Cell concentration: 600 cells/ml • Enrichment factor of 105
• Channel length: 6 and 13 cm