Skip to main content
. 2018 Feb 27;13(2):e0193023. doi: 10.1371/journal.pone.0193023

Table 1. Intraobserver repeatability of ocular parameters measured by Tomey OA-2000 biometer (n = 108).

Parameter Observer Mean ± SD CoV (%) Sw 2.77Sw ICC
AL (mm) 1st 24.56 ± 1.30 0.034 0.013 0.037 1.000
2nd 24.56 ± 1.30 0.035 0.011 0.030 1.000
CCT (μm) 1st 520.48 ± 31.63 0.668 5.08 14.06 0.991 (0.988–0.994)
2nd 521.05 ± 31.69 0.591 4.75 13.17 0.993 (0.990–0.995)
ACD (mm) 1st 3.59 ± 0.32 0.426 0.020 0.057 0.999 (0.998–0.999)
2nd 3.58 ± 0.32 0.385 0.016 0.045 0.999 (0.998–0.999)
LT (mm) 1st 3.72 ± 0.29 0.651 0.055 0.152 0.990 (0.987–0.993)
2nd 3.73 ± 0.30 0.686 0.063 0.175 0.987 (0.982–0.991)
Kf (D) 1st 42.81 ± 1.44 0.141 0.070 0.200 0.999 (0.999–0.999)
2nd 42.81 ± 1.41 0.140 0.077 0.214 0.999 (0.999–0.999)
Ks (D) 1st 43.84 ± 1.61 0.220 0.134 0.372 0.998 (0.997–0.998)
2nd 43.84 ± 1.63 0.183 0.134 0.372 0.998 (0.997–0.998)
Km (D) 1st 43.32 ± 1.48 0.174 0.055 0.152 1.000 (0.999–1.000)
2nd 43.32 ± 1.49 0.113 0.070 0.196 0.999 (0.999–1.000)
PD (mm) 1st 5.53 ± 0.88 5.298 0.363 1.001 0.943 (0.922–0.960)
2nd 5.49 ± 0.87 4.885 0.322 0.893 0.954 (0.936–0.967)
CD (mm) 1st 11.88 ± 0.39 0.829 0.197 0.547 0.916 (0.885–0.940)
2nd 11.87 ± 0.39 0.575 0.170 0.472 0.936 (0.912–0.954)

AL: axial length, CCT: central corneal thickness, ACD: anterior chamber depth, LT: lens thickness, Kf: flat keratometry, Ks: steep keratometry, Km: mean keratometry, PD: pupil diameter, CD: white to white, D: diopter, SD: standard deviation, CoV: coefficient of variance, Sw: within-subject standard deviation, ICC: intraclass correlation coefficient.