Table 2. Interobserver reproducibility of ocular parameters measured by Tomey OA-2000 biometer (n = 108).

Parameter CoV (%) Sw 2.77Sw ICC
AL (mm) 0.022 0.008 0.023 1.000 (1.000–1.000)
CCT (μm) 0.423 3.016 8.355 0.995 (0.993–0.997)
ACD (mm) 0.332 0.022 0.061 0.998 (0.997–0.998)
LT (mm) 0.550 0.058 0.160 0.981 (0.972–0.987)
Kf (D) 0.082 0.045 0.124 0.999 (0.999–0.999)
Ks (D) 0.110 0.073 0.202 0.999 (0.999–0.999)
Km (D) 0.068 0.045 0.124 0.999 (0.999–0.999)
PD (mm) 3.855 0.370 1.020 0.944 (0.918–0.962)
CD (mm) 0.378 0.090 0.248 0.973 (0.961–0.982)

AL: axial length, CCT: central corneal thickness, ACD: anterior chamber depth, LT: lens thickness, Kf: flat keratometry, Ks: steep keratometry, Km: mean keratometry, PD: pupil diameter, CD: white to white, D: diopter, SD: standard deviation, CoV: coefficient of variance, Sw: within-subject standard deviation, ICC: intraclass correlation coefficient.