Table 4. Comparison of ocular parameters measured by Tomey OA-2000 biometer and IOLMaster (n = 108).

Pairings Mean Difference ± SD 95% CI t Value p* Value
AL -0.058 ± 0.094 -0.076 to -0.040 -6.405 < 0.05
Kf -0.088 ± 0.150 -0.117 to -0.060 -6.107 < 0.05
Ks -0.163 ± 0.170 -0.195 to -0.130 -9.978 < 0.05
Km -0.127 ± 0.117 -0.150 to -0.105 -11.281 < 0.05
ACD 0.010 ± 0.075 -0.004 to 0.024 1.382 0.169
CD -0.171 ± 0.217 -0.213 to -0.130 -8.188 < 0.05

AL: axial length, CCT: central corneal thickness, ACD: anterior chamber depth, LT: lens thickness, Kf: flat keratometry, Ks: steep keratometry, Km: mean keratometry, PD: pupil diameter, CD: white to white, D: diopter, SD: standard deviation, CI: confidence interval,

*: two tailed.