. 2018 Feb 10;410(8):2095–2110. doi: 10.1007/s00216-017-0819-3

Table 1.

Intermediate precision of optimized CE-SDS assay

Parameter Mean ± uca CV
Instrument Qualification (IQ) Standard
 10 kDa internal standard migration time (min) 12.45 ± 0.12 1.0%
 100 kDa marker migration time (min) 21.97 ± 0.21 1.0%
PS 8670
 Monomer migration time (min)b 28.19 ± 0.28 1.0%
 Monomeric purity (%)b 98.79 ± 0.38 0.4%
 Light chain migration time (min)c 15.29 ± 0.18 1.2%
 Heavy chain migration time (min) c 19.29 ± 0.23 1.2%
 Light chain relative abundance (%) 32.02 ± 0.20 0.6%
 Heavy chain relative abundance (%) 67.26 ± 0.20 0.3%
 Glycan occupancy (%)c 99.40 ± 0.01 0.01%
 NRS relative abundance (%)c 0.31 ± 0.02 4.9%

aStated uncertainty represents the intermediate precision reported as a combined standard uncertainty, at a level of one standard deviation, based on ANOVA analysis as described in ESM(n = 12)

bnon-reduced sample

creduced sample