Table 2.
Parameter | Mean ± uca | CV |
---|---|---|
Instrument Qualification (IQ) Standard | ||
γ-globulin Migration Time (min) | 3.90 ± 0.08 | 2.0% |
Ovalbumin Migration Time (min) | 4.69 ± 0.10 | 2.0% |
PS 8670 | ||
Monomeric Purity (%) | 98.78 ± 0.12 | 0.1% |
HMW Relative Abundance (%) |
1.02 ± 0.12 | 11.6% |
LMW Relative Abundance (%) |
0.20 ± 0.01 | 4.1. % |
Monomer Retention Time (min) | 3.95 ± 0.08 | 2.1% |
Resolution (Dimer-Monomer) | 1.95 ± 0.05 | 2.7. % |
aStated uncertainty represents the intermediate precision reported as a combined standard uncertainty, at a level of one standard deviation, based on ANOVA analysis as described in ESM (n = 24)