TABLE II.

Summary of the measurement data obtained for the bone cements using the instrument. The variance in the data represents one standard deviation.

Palacos R
Compliance
(μm/N) DCa (%) PSa (MPa) PPEb (°C) tSETc (min)
0.33 77.1 ± 1.3 1.513 ± 0.087 3.75 ± 0.21 11.56 ± 0.86
0.79 77.3 ± 2.6 1.155 ± 0.056 3.17 ± 0.13 10.39 ± 0.99
4.21 80.4 ± 3.7 0.446 ± 0.038 2.41 ± 0.53 12.62 ± 0.43
12.27 78.2 ± 2.3 0.250 ± 0.030 2.33 ± 0.47 13.42 ± 0.49
26.96 78.0 ± 3.1 0.176 ± 0.020 3.23 ± 0.76 11.47 ± 0.91
50.27 76.7 ± 3.4 0.111 ± 0.009 3.30 ± 0.27 12.05 ± 0.40
84.18 78.0 ± 0.7 0.081 ± 0.003 2.96 ± 0.47 10.31 ± 1.30
Palacos LV
Compliance
(μm/N) DCa (%) PSa (MPa) PPEb (°C) tSETc (min)
0.33 78.4 ± 3.5 1.314 ± 0.023 2.30 ± 0.4 15.50 ± 0.42
0.79 80.2 ± 0.8 1.145 ± 0.032 2.60 ± 0.2 13.50 ± 0.16
4.21 81.2 ± 3.6 0.440 ± 0.025 3.00 ± 1.03 14.44 ± 1.95
12.27 83.5 ± 2.8 0.231 ± 0.020 3.46 ± 0.39 13.96 ± 1.60
26.96 80.6 ± 3.0 0.146 ± 0.008 2.57 ± 0.75 12.85 ± 3.61
50.27 82.0 ± 2.7 0.125 ± 0.008 3.10 ± 0.78 14.74 ± 0.77
84.18 80.1 ± 1.4 0.078 ± 0.008 3.15 ± 0.37 15.13 ± 0.46
a

Values at the end of 1 h of the measurement.

b

PPE—Peak of the PE.

c

tSET—Setting time, time taken to reach half of the PPE.