Table 3.

Individual laboratory relative potency estimates (geometric mean) for CDC assays.

  Versus IH Refa
Versus candidate (Sample A)
  Candidate (Sample A)
Sample C
Coded duplicate (Sample B)
Sample C
Lab
GM
%GCV
n
GM
%GCV
n
GM
%GCV
n
GM
%GCV
n
1 1.01 10 9 0.98 12 8 0.99 16 8 0.99 17 8
2 Not available     1.03 11 3 0.95 19 3
3 1.00 7 9 0.98 8 9 1.00 5 9 0.98 5 9
4 Not available     1.03 10 9 0.97 7 9
5 1.15 7 9 1.12 9 9 0.96 7 9 0.98 10 9
6 1.01 n/a 2 1.07 7 4 0.95 n/a 2 1.02 2 3
7 1.08 12 14 1.01 7 8 0.99 6 7 0.92 13 7
8 0.99 9 8 0.93 9 8 0.98 9 8 0.95 9 7
9 1.26 4 8 1.22 8 8 0.99 4 9 0.96 8 9
10 1.48 n/a 2 n/a n/a 0 1.05 6 5 1.16 n/a 2
11 1.11 12 9 1.13 13 8 1.02 10 8 1.03 10 8
12 1.42 n/a 2 1.47 n/a 1 1.00 5 4 0.97 6 3
13 0.96 7 9 0.95 4 9 1.00 9 9 0.99 6 9
14 Not available     0.98 4 8 0.98 6 7
15 0.99 4 9 0.99 3 9 1.00 3 9 1.00 3 9
16 0.95 6 8 0.94 7 8 1.00 7 7 0.99 4 8
a

“In house” reference standard.